Accuracy
Numbers from images the models never saw in training. Last updated 24 November 2026.
| Module | Test set | Result |
|---|---|---|
| Ceph landmarks (29 points) | 150 lateral cephs, Aariz test split | Mean radial error 1.12 mm; 86.4% of landmarks within 2 mm (best tested model, v4) |
| CVM staging | Aariz CVM labels | 64.7% exact stage; 96.7% within one stage |
| Dental age | Panoramic set, ages 9–19 | Average error 0.88 years |
| Panoramic findings | DentalOPG test split | In development, not yet in the report |
Limits you should know
- All tests so far use public research datasets. Local X-ray machines, exposure and patient groups can change accuracy, and local clinical validation is the next step.
- Some landmarks are harder than others; gonion, articulare and condylion carry the largest errors and deserve a manual check.
- Dental age is less reliable at the extremes of the range; below 11 and above 17 years the app reports "probably younger" or "probably older".
- Exact CVM stage is right about two times in three; the growth phase is cross-checked with age before it is shown.
Results are updated here when a new model version is released.